Mercury versus Dinamap blood pressure measurement: a randomised crossover trial
thesis
posted on 2017-02-08, 03:54authored byHeinemann, Megan
A routine aspect of nursing practice is obtaining accurate and reliable assessments of
patients' blood pressure. Traditionally, blood pressures have been measured using a
manual mercury sphygmomanometer. In recent years the trend has been to use
automated blood pressure machines. However, many nurses question the accuracy and
reliability of these machines when used in the clinical setting. A review of studies that
have evaluated the use of automated blood pressure machines provide conflicting results
across a variety of patient populations and clinical settings. Of note is that no studies
have been reported that test the use of automated blood pressure machines with adult
hospital patients presenting with a range of common illnesses and health concerns.
The purpose of this study was to determine the accuracy and reliability of a commonly
used automated blood pressure machine (Dinamap 81 00) relative to the manual mercury
sphygmomanometer. A randomized crossover design was used to test the hypothesis
that there would be no significant difference in systolic and diastolic blood pressure
readings between the two devices. Blood pressure measurements were obtained from a
sample of 63 adult in-patients recruited from medical, surgical, orthopaedic and critical
care wards of a large hospital in regional Victoria, Australia. Blood pressure
measurements for each patient were taken concurrently by two registered nurses using a
manual and an automated device, with type of device and the arm from which the
measures were obtained randomly determined. Each nurse was blind to the readings
obtained by the other nurse. To control for potential threats to the internal validity of the
study, the technique for taking blood pressure readings was standardized as was cuff
sizes. A data collection form was designed to record systolic and diastolic blood pressure
readings, type of device, cuff size, time and date of measurement, and patient details
such as age, gender and diagnosis.
To test the hypothesis of 'no difference' between blood pressures taken with the Dinamap
machine and those taken using a manual sphygmomanometer, level of agreement for
systolic and diastolic readings were examined using four sets of criteria: intercorrelations
between automated and manual readings, the Association for the
Advancement of Medical Instrumentation (AAMI) criteria, the British Hypertensive
Society (BHS) grading system, and Bland-Altman plots. Separate analyses were
conducted for the two assessors: Nurse A with extensive nursing experience and Nurse
B with less experience. Results of analyses for Nurse A supported the research
hypothesis on all four criteria for both systolic and diastolic blood pressure readings. For
Nurse B only partial support was provided, with agreement between the two devices on
systolic but not diastolic readings. Thus the recommendation from this study is that
nurses can use the Dinamap with confidence to assess blood pressures in a general
population of adult hospital in-patients, but with the proviso the nurse taking the
measures has sufficient experience. Comparison of mean differences in automated and
manual measures found the Dinamap machine consistently under-read both systolic and
diastolic blood pressures. With the exception of one diastolic comparison, these
differences were not considered to be clinically significant.
Findings from the study are discussed in relation to the research literature and
implications for clinical practice. Also considered are methodological issues and
directions for further investigation.